“Not detected” on a lab report: LOD, LOQ and limits

Read “not detected” with the analyte, method, reporting threshold and units. Understand detection and quantification before comparing results.

Public-source desk researchEditorial owner: Andrei SasContent reviewed

A “not detected” result needs a method and a threshold to be useful. It does not mean the laboratory has proved an absolute zero. Before comparing suppliers, establish what was tested, what the laboratory could detect and how it reports low results.

Detection and quantification answer different questions.

IUPAC defines the detection limit in relation to what a specified analytical procedure can detect with reasonable certainty. Its quantification-limit terminology concerns measured values that meet a defined fitness-for-purpose requirement. Detecting a signal and reporting a sufficiently reliable quantity are different tasks.

Laboratories may use “ND”, “below LOQ” or other qualifiers under their own reporting conventions. Read the report legend and request an explanation when necessary. Do not assume every “ND” label refers to the same threshold or calculation.

Keep the limit beside the result.

The following is our proposed extraction table for a procurement file. It helps prevent a result from losing its meaning when someone copies it into a spreadsheet or presentation. A result without the associated threshold should remain incomplete.

Low-result review fields
FieldQuestion to resolve
AnalyteWhich specific substance was included in the test?
MethodWhich procedure and sample preparation were used?
Result qualifierWhat does ND or another symbol mean here?
Applicable limitIs this the LOD, LOQ or a reporting threshold?
Units and basisConcentration, mass fraction or amount in the tested sample?
Sample identityWhich submitted material does the result describe?

Compare thresholds before comparing labels.

In a hypothetical example, one report states ND with a reporting threshold of 1 unit, while another states ND with a threshold of 10 units. Even if the units and methods are otherwise comparable, those two labels do not express the same sensitivity. Neither result establishes absolute absence.

Now suppose a third report provides no limit at all. Ranking it beside the others as equally reassuring hides missing information. Ask for the applicable limit and reporting convention. If the answer remains unavailable, write “threshold not established” in the comparison.

A targeted test also answers a targeted question. An undetected result for one named analyte does not establish that every possible impurity or contaminant was assessed. The test panel and sample identity belong beside the conclusion.

Use the laboratory’s requirement as the reference.

Before requesting a test, the scientific or quality owner should define the relevant analytes and decision requirements. That allows a laboratory to explain whether its procedure is suitable for the question. Choosing a method only after seeing an attractive result reverses the decision process.

For a supplier enquiry, request the full report, its result legend and the relevant limits. Ask the issuer to clarify ambiguous qualifiers. Keep this evidence with the batch record, and avoid translating a limited analytical finding into a general safety or human-use claim.